Table 2.

Ten selected QRS detection algorithms.

Methods Filtering Extracting features Setting threshold Postprocessing
Pan–Tompkins algorithm [15] 5–15 Hz band-pass filter Derivative; square; integrate Two sets of adaptive thresholds Searching back; T wave judging
Hamilton-mean algorithm [11]
Hamilton-median algorithm [11]
RS slope algorithm [2123] Median filter Derivative; detecting negative slope 10 groups of duration empirical thresholds; one fixed amplitude threshold 200 ms refractory blanking technology
Sixth power algorithm [24] Two-stage median filter Sixth power One adaptive threshold Determining end point K
Finite state machine (FSM) algorithm [25] / Derivative; integrate; square Three thresholding stages /
U3 transform algorithm (U3) [26] 8–30 Hz band-pass filter U3 transform Two fixed thresholds Searching back; 270 ms refractory blanking technology
Difference operation algorithm (DOM) [2, 27] 8–30 Hz band-pass filter Derivative; detecting positive extreme points Positive threshold; negative threshold Optimizing; matching filtered signal
“jqrs” algorithm [2830] Sombrero hat-like low-pass filter Integrate One fixed threshold Searching back; 200 ms refractory blanking technology
Optimized knowledge-based algorithm (OKB) [1] 8–20 Hz band-pass filter Squaring; integration Two dynamic thresholds Determining the maxima of each block as R peak