Table 4.
Maximum temperature in the thermal sensor next to the antenna slots for the computational model, phantom experimentation, and in vivo swine experimentation.
FEM model | Phantom experimentation | In vivo experimentation | FEM model versus phantom | FEM model versus in vivo | Phantom versus in vivo | |
---|---|---|---|---|---|---|
Power/time (W/s) | Maximum temperature next to the slots (°C) | ΔT (°C) | ||||
10/120 | 92.11 | 70.91 | 89.37 | 21.20 | 2.74 | 18.46 |
20/40 | 102.60 | 69.15 | 91.31 | 33.45 | 11.29 | 22.16 |
30/30 | 109.20 | 76.78 | 90.42 | 32.42 | 18.78 | 13.64 |
50/15 | 136.00 | 75.22 | 103.96 | 60.78 | 32.04 | 28.74 |
| ||||||
Power/time (W/s) | Maximum temperature at 1 cm (°C) | ΔT (°C) | ||||
10/120 | 69.35 | 63.60 | 72.88 | 5.75 | 3.53 | 9.28 |
20/40 | 70.40 | 59.84 | 64.61 | 10.56 | 5.79 | 4.77 |
30/30 | 84.85 | 67.11 | 69.45 | 17.74 | 15.40 | 2.34 |
50/15 | 80.60 | 63.71 | 85.27 | 16.89 | 4.67 | 21.56 |