. 2016 Nov 2;43(12):6301–6308. doi: 10.1118/1.4966032

Table I.

Peak and valley doses measured by film dosimetry at 1, 5, and 10 mm depth in a solid water phantom (Fig. 1). PVDRs in this domain are similar to PVDRs expected at the ESRF for a 20 × 20 mm2 microbeam field.

Depth (mm) 1 5 10
Peak (mGy/s) 300 ± 20 230 ± 10 170 ± 10
Valley (mGy/s) 10.2 ± 0.5 11.2 ± 0.5 10.7 ± 0.5
PVDR 30 ± 3 21 ± 2 15.5 ± 1.5