Table 2.

Bolded items 1.5T and 3T, Consensus result in columns for routine clinical/multi-centre trial. +, agree but no consensus; *, agree with consensus; **, strongly agree with consensus. Italics indicate round table consensus

Clinical Trial
Acquisition Protocol
Do not use intrinsic body coil as a receiver coil ** **
Axial from cranial vertex to mid-thigh ** **
Total acquisition time < 30 mins ** **
Free breathing
 In-plane acquisition pixel size ≥ 3 × 3 mm (matrix 128 × 128 FoV ~ 380 × 420) * *
 Minimum slice thickness = 5 mm * *
40–50 slices per bed station * *
Single shot EPI (Contiguous slices, interleaved) ** **
Parallel imaging (Reduction factor = 2)
 Fat suppression method = STIR (160 ms <TI < 180 ms) * **
Minimum number of b-values = 2
b-values = 50–100 s mm2, 800–1000 s mm2 * *
Diffusion weighted images: trace weighted + +
 Number of averages = 3 (more at higher b value if possible, dependent on grad encode scheme) + *
TE < 95 ms
TR ≥ 5×T1 of tissue of interest
Optimisation Protocol
Always perform initial optimisation on suitable test object: * **
  minimally diffusing, (or doped water) * **
  known ADC values, * **
  controlled temperature (e.g. iced water phantom) * **
  larger than total number of slices (per bed station) * **
Minimise the:
  Static field distortion * **
  eddy-current induced distortion * **
  ghosting ** **
  Optimise fat suppression on human volunteers * **
  Assess repeatability and reproducibility for multi-centre trials on human volunteers + **
Routine Quality Assurance
Routine QC measurements should be performed and recorded ** **
  Not necessary to test daily or weekly + +
  every 3 months * *
  after software/hardware upgrades ** **
  after repairs/maintenance * *
Routine Quality control measurements should be performed with a suitable test object:
  iced water + *
  sugar/(doped)water at room temperature + *
Routine Quality control tests
  ADC linearity in Z-direction across all slices in bed station + *
  Static field distortion + *
  Eddy-current induced distortion + *
  Ghosting + *

ADC, apparent diffusion coefficient; FOV, field of view; TE, echo time; TR, repetition time; QA, quality assurance; QC, quality control; STIR, short tau inversion-recovery.