Skip to main content
. 2018 May 24;18:49. doi: 10.1186/s12894-018-0362-3

Table 2.

Catheter decolonization results – polyhexanide 0.02% versus NaCl 0.9%

Catheter Swab Extraction Method Membrane Filtration Method
log with NaCl log with polyhexanide Difference NaCl –polyhexanide log with NaCl log with polyhexanide Difference NaCl –polyhexanide
1 2.34 2.48 −0.13 1.00 1.00 0.00
2 2.48 1.48 1.00 1.00 1.00 0.00
3 2.48 1.30 1.18 2.00 0.00 2.00
4 2.48 1.48 1.00 2.00 0.00 2.00
5 2.48 1.48 1.00 2.00 2.00 0.00
6 2.48 2.48 0.00 2.00 2.00 0.00
7 2.48 2.48 0.00 1.00 1.00 0.00
8 2.48 2.48 0.00 2.00 1.00 1.00
9 2.08 2.48 −0.40 2.00 0.00 2.00
10 1.90 2.48 −0.57 2.00 0.00 2.00
Mean 2.37 2.06 0.31a 1.70 0.80 0.90b
Standard deviation 0.196 0.514 0.628 0.458 0.748 0.943
No of catheters 10 10

aThe difference between the polyhexanide and NaCl groups is not statistically significant (p = 0.173) for the swab extraction method

bThe difference between the polyhexanide and NaCl groups is statistically significant (p = 0.034) for the liquid extraction method