ABSTRACT
We have previously demonstrated that culturing periprosthetic tissue in blood culture bottles (BCBs) improves sensitivity compared to conventional agar and broth culture methods for diagnosis of prosthetic joint infection (PJI). We have also shown that prosthesis sonication culture improves sensitivity compared to periprosthetic tissue culture using conventional agar and broth methods. The purpose of this study was to compare the diagnostic accuracy of tissue culture in BCBs (subsequently referred to as tissue culture) to prosthesis sonication culture (subsequently referred to as sonicate fluid culture). We studied 229 subjects who underwent arthroplasty revision or resection surgery between March 2016 and October 2017 at Mayo Clinic in Rochester, Minnesota. Using the Infectious Diseases Society of America (IDSA) PJI diagnostic criteria (omitting culture criteria) as the gold standard, the sensitivity of tissue culture was similar to that of the sonicate fluid culture (66.4% versus 73.1%, P = 0.07) but was significantly lower than that of the two tests combined (66.4% versus 76.9%, P < 0.001). Using Bayesian latent class modeling, which assumes no gold standard for PJI diagnosis, the sensitivity of tissue culture was slightly lower than that of sonicate fluid culture (86.3% versus 88.7%) and much lower than that of the two tests combined (86.3% versus 99.1%). In conclusion, tissue culture in BCBs reached sensitivity similar to that of prosthesis sonicate fluid culture for diagnosis of PJI, but the two tests combined had the highest sensitivity without compromising specificity. The combination of tissue culture in BCBs and sonicate fluid culture is recommended to achieve the highest level of microbiological diagnosis of PJI.
KEYWORDS: prosthetic joint infection, PJI, periprosthetic tissue culture, sonicate fluid culture, blood culture bottles
INTRODUCTION
Prosthetic joint infection (PJI) is a severe complication of joint arthroplasty, with an infection rate ranging from 0.88% to 2.18% (1–3). PJI is associated with mortality, prolonged hospital stays, and high cost (2, 4). Accurate diagnosis is important in PJI disease management; however, diagnosis remains challenging because signs and symptoms of PJI are often subtle (5) and there is no diagnostic test with perfect accuracy (6). Periprosthetic tissue (subsequently referred to as tissue) culture results are important criteria for diagnosis of PJI as defined by the Infectious Diseases Society of America (IDSA) and the Musculoskeletal Infection Society (MSIS) (6, 7). Currently, conventional tissue culture using agar and thioglycolate broth is the most commonly used clinical microbiological test. However, it has been reported as having low sensitivity, ranging from 33% for agar cultures to 44% with combined agar and thioglycolate broth culture when using IDSA PJI criteria (8).
With the recognition of the role of bacterial biofilms on prosthesis surfaces, Tunney et al. first applied sonication to dislodge adherent bacteria from explanted prosthetic hips (9). Our group further developed a clinically useful implant sonication technique and demonstrated that using sonication to dislodge adherent bacteria for culture yielded higher sensitivity than conventional tissue culture for diagnosis of prosthetic hip and knee infection (78.5% versus 60.8%, P < 0.001) (10). Subsequently, we showed that the sonication method had higher sensitivity than conventional tissue culture in shoulder arthroplasty infection (66.7% versus 54.5%, P = 0.046) and spinal implant infection (91% versus 73%, P = 0.046) (11, 12). Furthermore, several other groups have demonstrated that the sonication method improves the diagnosis of orthopedic implant-associated infection (13–16).
Despite the high sensitivity of sonicate fluid culture for diagnosis of PJI, periprosthetic tissue remains the predominant sample type used for culture-based diagnosis of PJI. Recently, we compared culturing tissue in blood culture bottles (BCBs) to conventional agar and thioglycolate broth cultures for diagnosis of PJI using Bayesian latent class modeling (LCM), a statistical method that can be used to define diagnostic accuracy in the absence of a “gold standard” (17). We demonstrated culturing tissue in BCBs to be more sensitive than agar and thioglycolate broth cultures. Tissue culture in BCBs, as well as sonicate fluid culture, is now standard practice at our institution (8). However, the sensitivity of tissue culture in BCBs compared to that of prosthesis sonicate fluid culture is unknown. In particular, it is not clear whether culturing periprosthetic tissue specimens in BCBs reaches the diagnostic accuracy of prosthesis sonication culture.
Therefore, the purpose of this study was to compare the diagnostic accuracy of periprosthetic tissue culture in BCBs to that of prosthesis sonicate fluid culture for the diagnosis of PJI. We applied Bayesian LCM for our analysis, in addition to using standard nonmicrobiological IDSA PJI criteria for classification of our cases.
MATERIALS AND METHODS
Study design.
This study was a prospective cohort study of consecutive patients undergoing revision or resection hip, knee, shoulder, or elbow arthroplasty between March 2016 and September 2017 at Mayo Clinic, in Rochester, MN. This study was approved by the Mayo Clinic Inquisitional Review Board. Patients were excluded if they did not provide authorization of use of their health records (Minnesota Statute 144.335), no prosthesis was submitted for sonication culture, an antibiotic-loaded cement spacer was submitted for sonication culture, or fewer than two periprosthetic tissue specimens were cultured.
Patient classification.
Patients were classified as having PJI based on nonmicrobiological IDSA criteria when they had at least one of the following criteria: a sinus tract communicating with the prosthesis was present, acute inflammation was noted on histopathologic examination of periprosthetic tissue obtained during surgery, or purulence surrounding the prosthesis was documented by the surgeon (7). Microbiology culture results were not used to define PJI because diagnostic accuracy of culture-based testing was being evaluated. Aseptic failure was defined as cases not meeting criteria for PJI. Patient laboratory values for comparison were chosen based on the threshold for PJI diagnostic criteria by the Musculoskeletal Infection Society (18). Patients were also separated into a lower extremity (i.e., hip and knee) subgroup and an upper extremity (i.e., shoulder and elbow) subgroup for diagnostic accuracy analyses.
Microbiological methods.
Periprosthetic tissue culture was processed as previously described (8). Briefly, periprosthetic tissue collected during surgery was placed into a sterile vial containing CO2 (to maintain anaerobic conditions) and delivered to the clinical microbiology laboratory. Tissue specimens were homogenized for 1 min in 5 ml brain heart infusion broth (Seward Stomacher 80 Biomaster; Seward Inc., Port St. Lucie, FL), and then 1 ml of homogenized liquid was inoculated (using a 3-ml syringe with a 22-gauge [22G] needle) into each of a Bactec Plus Aerobic/F bottle (with resin) and a Bactec Lytic/10 Anaerobic/F bottle (without resin) (BD Diagnostic Systems) according to standard procedures at our institution. Bottles were incubated in a Bactec FX instrument (BD Diagnostic Systems) for 14 days. Positive bottles were subcultured to agar plates based on the results of Gram staining per standard laboratory practice. Periprosthetic tissue cultures were defined as positive when an identical microorganism was isolated from two or more tissue specimens or if Staphylococcus aureus, Enterococcus species, or yeast was isolated from a single tissue specimen. Sonication was performed as previously described (11). In brief, prostheses were collected and placed in sterile 1-liter straight-side wide-mouth polypropylene containers (Nalgene, Lima, OH). Sterile Ringer's solution (400 ml) was added to each container. The container was vortexed for 30 s and then sonicated for 5 min (40 kHz) using an ultrasonic cleaning bath (Branson 5510; Sonics, Richmond, VA), followed by additional vortexing for 30 s. The sonicate fluid was centrifuged at 4,000 rpm (3,150 × g) for 5 min and concentrated 1:100 to 4 ml; 0.1 ml of the concentrated sonicate fluid was inoculated onto aerobic sheep blood and chocolate agar and incubated in 5% CO2 for 5 days and onto anaerobic sheep blood agar and incubated anaerobically for 14 days. A single colony growing on a plate is equivalent to one colony per 10 ml sonicate fluid. A cutoff value of ≥20 CFU/10 ml was defined as positive for sonicate fluid cultures, in accordance with the previously established standard at our institution (11). Sonicate fluid culture was also considered positive if any S. aureus, Enterococcus species, or yeast growth was present. Organism identification was performed using routine laboratory methods, most commonly, matrix-assisted laser desorption ionization–time of flight mass spectrometry, using a Bruker Daltonics system (Billerica, MA).
Statistical analysis.
Descriptive summaries were reported as medians and interquartile ranges (IQRs) for continuous variables and as frequencies and percentages for categorical variables. Due to lack of a proper gold standard, the Bayesian LCM approach was used for the analysis. This approach overcomes potential flaws of traditional analysis and is based on the assumptions that no gold standard exists and that the true disease prevalence requires estimation, both relevant to PJI (17). We used this approach in our prior research work (8, 19). Using the Bayesian LCM Joseph et al. model (17), estimates of prevalence, sensitivity, and specificity were reported along with 95% credible intervals. We used uniform prior distribution for the prevalence of each latent variable as done in our prior study (8). Input data for Bayesian LCM are shown in Table S1 in the supplemental material. Further analysis was also performed using nonmicrobiological IDSA criteria as a gold standard (7). Sensitivity, specificity, NPV, and PPV were estimated, along with 95% exact binomial confidence intervals. Comparison of sensitivities and specificities among culture-types and the combination of tissue and sonicate fluid culture were performed in a pairwise manner using McNemar's test. P values of less than 0.05 were considered statistically significant. Analyses were performed using SAS version 9.4 (SAS Inc., Cary, NC) and Bayesian LCM software version 1.14 (20), which is available at http://www.nandinidendukuri.com/software/blcm.
RESULTS
Patient characteristics.
A total of 305 subjects undergoing revision or resection arthroplasties during the study period had prostheses sent for sonication culture. Seventy-six subjects were excluded: 24 did not have written authorization for use of their medical records, 37 had an antimicrobial spacer submitted for culture, and 15 had fewer than two periprosthetic tissue samples submitted for culture. The remaining 229 subjects were studied, and among them, 104 had PJI and 125 had aseptic failure according to IDSA nonmicrobiological PJI criteria by meeting at least one of the following criteria: a sinus tract communicating with the prosthesis, acute inflammation noted on histopathologic examination of periprosthetic tissue obtained during surgery, or purulence surrounding the prosthesis documented by the surgeon (7). Among 104 PJI subjects, 8.7% (9/104) met all three criteria, 38.4% (40/104) met two of three criteria, and 52.9% (55/104) met a single criterion. Demographic and clinical characteristics of the study subjects are shown in Table 1. Subject age and gender distributions and reasons for primary arthroplasty were similar between the two groups. Most subjects had pain without significant difference between the two groups. Fever and local signs of infection were, not unexpectedly, higher among PJI than aseptic failure subjects. PJI subjects more frequently had an elevated preoperative blood leukocyte count, C-reactive protein concentration, erythrocyte sedimentation rate, or synovial fluid leukocyte count, as well as positive synovial fluid cultures than did aseptic failure subjects (P < 0.0001). Eighteen percent of PJI subjects had revision or resection surgeries less than 3 months after prosthesis implantation compared to only 1% of the aseptic failure subjects (P < 0.0001). Eighty-four percent of aseptic failure subjects had surgeries performed >12 months after prosthesis implantation, compared to 57% of the PJI subjects (P < 0.0001).
TABLE 1.
Characteristic | Valuea for subjects with: |
Pb | |
---|---|---|---|
Prosthetic joint infection (n = 104) | Aseptic failure (n = 125) | ||
Median age (IQR), yr] | 66.5 (25, 92) | 67 (26, 87) | 0.825 |
Male gender | 61 (59) | 66 (53) | 0.375 |
Type of prosthetic joint | 0.013 | ||
Hip | 31 (30) | 16 (13) | |
Knee | 55 (53) | 85 (68) | |
Shoulder | 11 (10) | 12 (10) | |
Elbow | 7 (7) | 12 (10) | |
Reason for primary arthroplasty | 0.618 | ||
Osteoarthritis | 66 (64) | 86 (69) | |
Fracture or trauma | 23 (22) | 28 (22) | |
Rheumatoid arthritis | 5 (5) | 5 (4) | |
Avascular necrosis | 0 (0) | 1 (1) | |
Fibromyalgia | 1 (1) | 1 (1) | |
Tumor | 3 (3) | 2 (2) | |
Unknown | 6 (6) | 2 (2) | |
Presence of sinus tract | 23 (22) | 0 (0) | <0.0001 |
Visible purulence at implant site | 79 (76) | 0 (0) | <0.0001 |
Acute inflammation in periprosthetic tissue [no./total (%)] | 55/86 (64) | 0/117 (0) | <0.0001 |
Presence of clinical symptoms or signs | |||
Pain | 92 (89) | 119 (95) | 0.059 |
Local signs of infectionc | 48 (46) | 4 (3) | <0.0001 |
Fever | 8 (8) | 2 (2) | 0.025 |
Radiolucent lines present | 33 (32) | 49 (39) | 0.241 |
Preoperative laboratory findings [no./total (%)]d | |||
Blood leukocyte count >10 × 109/liter | 21/97 (22) | 15/110 (14) | <0.0001 |
Erythrocyte sedimentation rate >30 mm/h | 58/98 (59) | 16/116 (14) | <0.0001 |
Serum C-reactive protein concn >10 mg/liter | 72/100 (72) | 22/116 (19) | <0.0001 |
Synovial fluid leukocyte count >3 × 109/liter | 38/50 (76) | 14/56 (25) | <0.0001 |
Synovial fluid neutrophils >80% | 31/50 (62) | 6/56 (11) | <0.0001 |
Positive synovial fluid culture | 27/57 (48) | 2/63 (3) | <0.0001 |
Antibiotics received 4 weeks prior to surgery | 38 (37) | 10 (8) | <0.0001 |
Surgical procedure | <0.0001 | ||
Revision (including one-stage exchange) | 26 (25) | 102 (82) | |
Resection (including insertion of a spacer) | 78 (75) | 23 (18) | |
No. of tissue cultures [median (IQR)] | 4 (2, 10) | 3 (2, 10) | <0.0001 |
Timing from prosthesis implanted to surgery (mo) | <0.0001 | ||
<3 | 19 (18) | 1 (1) | |
3–12 | 26 (25) | 19 (15) | |
>12 | 59 (57) | 105 (84) |
Values are no. or no./total (%) unless interquartile range (IQR) is specified.
Age as a continuous variable was compared using the Wilcoxon rank sum test, and categorical variables were compared using Fisher's exact test.
Local signs of infection included redness and swelling around the surgical site.
Blood leukocyte count was the result within 1 week preoperative; other laboratory findings were the latest results within 6 months.
Microbiology.
In total, 85 of 229 subjects had microorganisms detected by tissue culture or sonicate fluid culture, and the majority were monomicrobial (74 subjects, 87.1%), with a small portion being polymicrobial (11 subjects, 12.9%). Staphylococcus aureus (n = 17) and Staphylococcus epidermidis (n = 15) were the most prevalent organisms among PJI subjects; S. aureus was the predominant pathogen among prosthetic hip infections, and S. epidermidis was the most common organism in prosthetic knee and elbow infections. Cutibacterium acnes (n = 11) was the third most commonly identified organism and was the dominant organism in prosthetic shoulder infection. It took 3 to 13 days for C. acnes to grow in blood culture bottles and 5 to 12 days for C. acnes to grow in sonicate fluid culture. An unusual finding was that one PJI subject had Mycobacterium bovis BCG detected through periprosthetic tissue mycobacterial culture, while the aerobic and anaerobic cultures of periprosthetic tissue and sonicate fluid were both negative. Microbiological findings in periprosthetic tissue and sonicate fluid culture are shown in Table 2. Sixty-five (62.5%) PJI subjects had microorganisms detected by both tissue culture and sonicate fluid culture. Fifteen (14.4%) PJI subjects had organisms detected only by tissue culture or sonicate fluid culture, and the remaining 24 (23.1%) PJI subjects had no microorganism detected with either method. One hundred twenty (96.0%) subjects with aseptic failure had negative results from both tests. However, five (4.0%) subjects with aseptic failure had tissue cultures defined as positive with an identical microorganism isolated from two or more tissue specimens. Four of these tissue cultures had C. acnes, and the corresponding sonicate fluid cultures had less than 20 CFU/10 ml organisms isolated (without further identification performed), which was below the threshold of positivity. The fifth subject had positive tissue cultures for Staphylococcus capitis without any organism growth in sonicate fluid.
TABLE 2.
Category and no. of patients and culture results | Microorganism(s) (no. of patients) | No. of affected prosthetic joints |
|||
---|---|---|---|---|---|
Hip | Knee | Shoulder | Elbow | ||
Patients with prosthetic joint infection (104) | |||||
Positive tissue and sonicate fluid cultures (65) | 20 | 33 | 7 | 5 | |
Concordant (58) | |||||
Monomicrobial (54) | Staphylococcus aureus (15) | 8 | 6 | 1 | |
Staphylococcus epidermidis (11) | 1 | 9 | 1 | ||
Streptococcus mitis group (5) | 5 | ||||
Cutibacterium acnes (5) | 5 | ||||
Staphylococcus lugdunensis (3) | 1 | 2 | |||
Enterococcus faecalis (3) | 1 | 1 | 1 | ||
Staphylococcus caprae (1) | 1 | ||||
Staphylococcus simulans (1) | 1 | ||||
Streptococcus agalactiae (1) | 1 | ||||
Streptococcus dysgalactiae (1) | 1 | ||||
Streptococcus salivarius group (1) | 1 | ||||
Clostridium ramosum (1) | 1 | ||||
Corynebacterium striatum (1) | 1 | ||||
Escherichia coli (1) | 1 | ||||
Lelliottia amnigena (1) | 1 | ||||
Pseudomonas aeruginosa (1) | 1 | ||||
Enterobacter cloacae (1) | 1 | ||||
Prevotella bivia (1) | 1 | ||||
Polymicrobial (4) | S. epidermidis + Staphylococcus pseudintermedius (1) | 1 | |||
S. mitis + E. cloacae + Finegoldia magna (1) | 1 | ||||
S. aureus + P. aeruginosa (1) | 1 | ||||
Candia albicans + E. faecalis (1) | 1 | ||||
Discordant (7) (organisms detected in tissue/sonicate fluid) | E. cloacae (1)/E. cloacae + Finegoldia magna (1) | 1 | |||
S. epidermidis (1)/S. epidermidis + Corynebacterium amycolatum (1) | 1 | ||||
Aerococcus sanguinicola + P. aeruginosa (1)/A. sanguinicola + Anaerococcus murdochii (1) | 1 | ||||
S. aureus + E. faecalis + E. coli (1)/E. faecalis (1) | 1 | ||||
Candida parapsilosis + C. acnes (1)/C. parapsilosis (1) | 1 | ||||
Staphylococcus saccharolyticus + C. acnes (1)/C. acnes + Enterococcus faecium (1) | 1 | ||||
S. lugdunensis (1)/S. lugdunensis + E. coli (1) | 1 | ||||
Positive tissue and negative sonicate fluid cultures (4) | 2 | 1 | 1 | ||
C. acnes (1) | 1 | ||||
S. epidermidis (1) | 1 | ||||
S. aureus (1) | 1 | ||||
S. aureus + C. striatum (1) | 1 | ||||
Negative tissue and positive sonicate fluid cultures (11) | 5 | 2 | 2 | 2 | |
S. epidermidis (3) | 1 | 1 | 1 | ||
S. capitis (2) | 2 | ||||
S. caprae (1) | 1 | ||||
S. aureus (1) | 1 | ||||
S. mitis group (1) | 1 | ||||
C. striatum (1) | 1 | ||||
C. acnes (1) | 1 | ||||
Parvimonas micra (1) | 1 | ||||
Negative tissue and sonicate fluid cultures (24) | 4 | 19 | 1 | ||
Patients with aseptic failure (125) | |||||
Positive tissue and negative sonicate fluid cultures (5) | 2 | 3 | |||
C. acnes (4) | 1 | 3 | |||
S. capitis (1) | 1 | ||||
Negative tissue and sonicate fluid cultures (120) | 16 | 83 | 9 | 12 |
Sensitivity and specificity.
Using IDSA nonmicrobiological criteria as the gold standard for diagnosing PJI, the sensitivity of tissue culture was not statistically different from that of sonicate fluid culture (66.4%, 69/104 versus 73.1%, 76/104; P = 0.07), but was significantly lower than that of the two tests combined (66.4%, 69/104 versus 76.9%, 80/104; P < 0.001). The sensitivity of sonicate fluid culture alone was slightly lower than that of the two tests combined (73.1%, 76/104 versus 76.9%, 80/104; P = 0.045). The specificity of tissue culture was lower than that of sonicate fluid (96.0%, 120/125 versus 100%, 125/125; P = 0.025) and the same as that of the two tests combined (96.0%, 120/125 for both). In the lower extremity subgroup (i.e., hips and knees), the sensitivity of tissue culture was not different from that of sonicate fluid culture (65.1%, 56/86 versus 69.8%, 60/86; P = 0.205). In the upper extremity subgroup (i.e., shoulders and elbows), tissue culture showed lower sensitivity than sonicate fluid culture (72.2%, 13/18, versus 88.9%, 16/18), although the difference was not statistically significant (P = 0.179). Individual joint types were not analyzed due to the sample size.
When using Bayesian LCM analysis, the sensitivity of tissue culture was slightly lower than that of sonicate fluid culture (86.3% versus 88.7%), while the specificities were the same. In the lower extremity subgroup, the sensitivity of tissue culture was also slightly lower than that of sonicate fluid culture (88.4% versus 91.4%). In the upper extremity subgroup, tissue culture showed a trend toward lower sensitivity than sonicate fluid culture (68.4% versus 80.7%), with the combination of the two tests having the highest sensitivity, without compromising the specificity (Table 3).
TABLE 3.
Specimen or culture | Patients with PJI | Patients with aseptic failure | IDSA PJI criteria as gold standardb |
Bayesian latent-class modelingb |
|||||
---|---|---|---|---|---|---|---|---|---|
Sensitivity | Specificity | PPV | NPV | Prevalence | Sensitivity | Specificity | |||
All joints | 104 | 125 | 37.2 (31.1, 43.5) | ||||||
Tissue culture | 69 | 5 | 66.4 (56.4, 75.3)* | 96.0 (90.9, 98.7) | 93.2 (84.9, 97.8) | 77.4 (70.0, 83.7) | 86.3 (78.3, 92.4) | 99.6 (97.7, 100.0) | |
Sonicate fluid culture | 76 | 0 | 73.1 (63.5, 81.3)* | 100.0 (97.1, 100.0) | 100.0 (95.3, 100.0) | 81.7 (74.7, 87.5) | 88.7 (81.0, 94.3) | 99.6 (97.7, 100.0) | |
Combinationc | 80 | 5 | 76.9 (67.6, 84.6)* | 96.0 (90.9, 98.7) | 94.1 (86.8, 98.1) | 83.3 (76.2, 89.0) | 99.1 (95.7, 100.0) | 99.5 (97.6, 100.0) | |
Hips and knees | 86 | 101 | 34.8 (28.4, 41.8) | ||||||
Tissue culture | 56 | 2 | 65.1 (54.1, 75.1)* | 98.0 (93.0, 99.8) | 96.6 (88.1, 99.6) | 76.7 (68.5, 83.7) | 88.4 (79.6, 94.7) | 99.5 (97.1, 100.0) | |
Sonicate fluid culture | 60 | 0 | 69.8 (58.9, 79.2) | 100.0 (96.4, 100.0) | 100.0 (95.3, 100.0) | 79.5 (71.5, 86.2) | 91.4 (83.3, 96.5) | 99.4 (97.2, 100.0) | |
Combination | 63 | 2 | 73.3 (62.6, 82.2)* | 98.0 (93.0, 99.8) | 96.9 (89.3, 99.6) | 81.2 (73.1, 87.7) | 98.9 (94.7, 100.0) | 99.4 (96.9, 100.0) | |
Shoulders and elbows | 18 | 24 | 50.9 (36.1, 65.5) | ||||||
Tissue culture | 13 | 3 | 72.2 (46.5, 90.3)* | 87.5 (67.6, 97.3) | 81.3 (54.4, 96.0) | 80.8 (60.7, 93.5) | 68.4 (48.2, 84.3) | 96.8 (84.3, 99.9) | |
Sonicate fluid culture | 16 | 0 | 88.9 (65.3, 98.6) | 100.0 (85.8, 100.0) | 100.0 (79.4, 100.0) | 92.3 (74.9, 99.1) | 80.7 (62.8, 93.1) | 96.8 (84.7, 99.9) | |
Combination | 17 | 3 | 94.4 (72.7, 99.9)* | 87.5 (67.6, 97.3) | 85.0 (62.1, 96.8) | 95.5 (77.2, 99.9) | 97.0 (84.8, 99.9) | 96.6 (82.8, 99.9) |
PPV, positive predictive value; NPV, negative predictive value; *, statistically significant difference from other tests in the same comparison group, P < 0.05.
Values are % (95% confidence interval) for IDSA PJI and % (95% credible interval) for Bayesian latent-class modeling.
Combination includes tissue culture and sonicate fluid culture.
Effect of preoperative antimicrobial therapy on culture sensitivity.
Among 104 PJI subjects, 38 had received antimicrobial therapy within 4 weeks before surgery and 66 had not. Thirty-one of 80 PJI subjects (38.7%) with positive culture results had received antimicrobial therapy within 4 weeks before surgery, compared to 7 of 24 PJI subjects (23.1%) with negative culture results. The sensitivity of tissue culture in PJI subjects who had received antimicrobial therapy within 4 weeks before surgery was not different from that of PJI subjects who had not (71.1% versus 63.6%, P = 0.44). The sensitivity of sonicate fluid culture also showed no significant difference between those two groups (76.3% versus 71.2%, P = 0.57) (Table 4).
TABLE 4.
Culture type | No. (%) of subjects with positive culture |
Pa | |
---|---|---|---|
Previous antimicrobial therapy (n = 38) | No previous antimicrobial therapy (n = 66) | ||
Tissue culture | 27 (71.1) | 42 (63.6) | 0.44 |
Sonicate fluid culture | 29 (76.3) | 47 (71.2) | 0.57 |
Pb | 0.41 | 0.10 |
P value for comparison of sensitivity of each culture method alone in subjects who received and those who did not receive antimicrobial therapy.
P value for comparison of sensitivity between tissue culture and sonicate fluid culture in the subgroups of subjects who received or those who did not receive antimicrobial therapy.
DISCUSSION
In this study, we compared the diagnostic accuracy of periprosthetic tissue culture in BCBs to that of sonicate fluid culture for diagnosis of PJI. Two approaches were used. First, we used IDSA nonmicrobiological PJI criteria as the “gold standard” for classifying our subjects as having PJI. Some studies have included microbiological results as a criterion for diagnosis of PJI when evaluating microbiological methods (8, 21), whereas others have not (10, 22). We chose to use the nonmicrobiological approach, as including microbiological criteria would have de facto yielded 100% specificity of tissue culture. However, we recognize that this may have resulted in subject misclassification. Five aseptic failure subjects would have been classified as having PJI had microbiological criteria been used for subject classification. To overcome the lack of a perfect “gold standard,” and also because tissue and sonicate fluid cultures are conditionally independent, we compared test performances using Bayesian LCM analysis. Bayesian LCM is used to define diagnostic accuracy in the absence of a “gold standard” (17). It has been used to define diagnostic accuracy in infectious diseases, including typhoid fever, malaria, and pertussis (23–25), as well as, in our prior studies, PJI (8, 19). In both analyses, we found that tissue culture in BCBs reached sensitivity similar to that of sonicate fluid culture and that the combination of the two tests returned the best sensitivity. Notably, we have recently shown culture of periprosthetic tissue in BCBs to be cost saving compared to conventional (i.e., plate and broth) periprosthetic tissue culture methods (26).
Compared to conventional methods of tissue culture, culturing tissue in BCBs improves the performance of this method, reaching the performance of sonicate fluid culture. There are several possible reasons for this improvement. First, the resins in the Bactec Plus Aerobic/F bottle (BD Diagnostic Systems) may have bound antibiotics, neutralizing their effects and enhancing recovery of bacteria for subjects with previous antimicrobial therapy (27). Second, the inoculum volume into BCBs is large (1 ml into each bottle; 2 ml in total) compared to the volume inoculated onto agar plates (0.1 ml) and into thioglycolate broth (1 ml) using conventional tissue culture methods (8). The inoculum volume used for BCBs derived from a study by Hughes et al. (28) and has become our institutional standard. Third, culture conditions, including shaking, may increase sensitivity (29). Many studies have demonstrated that using BCBs can increase microbiological diagnostic sensitivity of PJI with various types of specimens, including synovial fluid, periprosthetic tissue, and even sonicate fluid (28–30). In addition, multiple tissue specimens cultured per patient (median, 4 for PJI subjects and 3 for aseptic failure subjects in this study) increase the chance of detecting microorganisms. Culturing tissue in BCBs also provides partial automation of the work flow and provides faster results than conventional tissue cultures (8, 22, 26), as addressed in our previous study (26). However, contamination of blood culture bottles is always a concern. It is not possible to enumerate the microorganisms grown in BCBs. Multiple tissue specimens are required; an identical microorganism isolated from two or more BCBs is defined as a positive result (19). Processing multiple tissue specimens increases the workload and takes a longer time than handling one prosthesis sonicate fluid culture (21). Interestingly, using Bayesian LCM analysis, tissue culture had the trend of lower sensitivity than sonicate fluid culture in the shoulder and elbow subgroup (68.4% versus 80.7%), with the same trend noted when using IDSA nonmicrobiological PJI criteria (72.2% versus 88.9%, P = 0.179).
In this study, we isolated multiple organisms from 11 of 85 (12.9%) subjects. This is similar to findings in previous studies (10, 13). In agreement with a previous study (8), S. aureus, followed by S. epidermidis, was the most common organism isolated from PJI subjects. C. acnes was the dominant pathogen in subjects with prosthetic shoulder infections, as previously reported (11), which also demonstrated that using anaerobic BCBs as well as aerobic ones for tissue culture is important, especially in shoulder PJIs.
Some studies suggest that preoperative antimicrobial therapy differentially affects the sensitivity of tissue and sonicate fluid culture (10, 31), while others have not corroborated this (32, 33). In this study, we found that preoperative antimicrobial therapy had no effect on culture sensitivities of tissues in BCBs or sonicate fluid. For periprosthetic tissue culture, the reason may be the resin in the Bactec Plus Aerobic/F bottle. Additionally or alternatively, the liquid medium in the BCBs may dilute out any antibiotic present. For sonicate fluid culture, a possible explanation could be that the Ringer's solution added for sonication was removed after centrifugation, diluting out antibiotics. This is supported by results of a study demonstrating that concentration of orthopedic implant sonicate fluid through centrifugation yielded higher culture sensitivity than did membrane-filtered sonicate fluid (34). In our previous study, we concluded that antimicrobial use reduced the sensitivity of periprosthetic tissue and sonicate fluid cultures (10); however, periprosthetic tissue specimens were not placed into BCBs and sonicate fluid was not concentrated in that study. In a study by Portillo et al. in 2013, the authors reported that antimicrobial therapy reduced the sensitivity of unconcentrated sonicate fluid culture (31). In a later study by the same group, it was demonstrated that when inoculating sonicate fluid into BCBs, the sensitivity was not affected by antimicrobial treatment (35). Overall, these results suggest that either of the methods studied here have the potential to overcome some effects of prior antimicrobial therapy.
In this study, 23.1% (24/104) of PJI subjects (defined by nonmicrobiological IDSA diagnostic criteria) had culture-negative infections, despite having tissue cultures in BCBs and sonicate fluid cultures performed. A possible explanation could be that they were infected by unusual microorganisms (e.g., mycobacteria, fungi) that were not detected by the culture methods under evaluation and would need specialized culture methods (36, 37). One PJI subject in this study had M. bovis BCG detected through periprosthetic tissue mycobacterial culture, while the aerobic and anaerobic cultures of periprosthetic tissue and sonicate fluid were both negative. Notably, not all subjects had mycobacterial or fungal cultures performed. Tools other than culture, such as organism-specific PCR or targeted or shotgun metagenomic sequencing, may be needed to detect pathogens (e.g., Ureaplasma species, Mycoplasma species) undetected by conventional culture (38, 39). Another possible reason could be that some of these subjects were misclassified as having infection.
There are several limitations in our study. First, we have used nonmicrobiological IDSA criteria to classify PJI. This could lead to misclassifying PJI patients with positive culture results but not meeting any of the other two criteria into the aseptic failure group. However, using Bayesian latent class modeling allows us to estimate diagnostic performance in the absence of a “gold standard.” Second, we defined tissue culture positivity as the identical microorganism detected in two tissue specimens and sonicate fluid culture positivity as growth of ≥20 CFU/10 ml (with some exceptions). The cutoff values chosen may underestimate the sensitivity of the evaluated tests. However, cutoff values are important for distinguishing pathogens from contaminants (8, 10). Third, this study did not include mycobacterial and fungal cultures. Finally, only the Bactec blood culture system and specific types of Bactec BCBs were evaluated.
In conclusion, periprosthetic tissue culture in BCBs reaches sensitivity similar to that of prosthesis sonicate fluid culture for diagnosis of PJI, but the combination of those two tests has the highest sensitivity without compromising the specificity. The combination of tissue culture in BCBs and sonicate fluid culture is recommended to achieve the highest level of microbiological diagnosis of PJI.
Supplementary Material
ACKNOWLEDGMENTS
This work was supported by the National Institute of Arthritis and Musculoskeletal and Skin Diseases, National Institutes of Health (R01 AR056647). R.P. is also supported by the National Institute of Allergy and Infectious Diseases, National Institutes of Health (R01 AR056647, R21 AI125870).
R.P. reports grants from CD Diagnostics, BioFire, Curetis, Merck, Hutchison Biofilm Medical Solutions, Accelerate Diagnostics, Allergan, and The Medicines Company. R.P. is or has been a consultant to Curetis, Specific Technologies, Selux Dx, GenMark Diagnostics, PathoQuest, and Genentech; monies are paid to Mayo Clinic. In addition, R.P. has a patent on a Bordetella pertussis/B. parapertussis PCR assay issued, a patent on a device/method for sonication, with royalties paid by Samsung to Mayo Clinic, and a patent on an antibiofilm substance issued.
Footnotes
Supplemental material for this article may be found at https://doi.org/10.1128/JCM.00319-18.
REFERENCES
- 1.Tande AJ, Patel R. 2014. Prosthetic joint infection. Clin Microbiol Rev 27:302–345. doi: 10.1128/CMR.00111-13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Kurtz SM, Lau E, Schmier J, Ong KL, Zhao K, Parvizi J. 2008. Infection burden for hip and knee arthroplasty in the United States. J Arthroplasty 23:984–991. doi: 10.1016/j.arth.2007.10.017. [DOI] [PubMed] [Google Scholar]
- 3.Marang-van de Mheen PJ, Bragan Turner E, Liew S, Mutalima N, Tran T, Rasmussen S, Nelissen R, Gordon A. 2017. Variation in prosthetic joint infection and treatment strategies during 4.5 years of follow-up after primary joint arthroplasty using administrative data of 41397 patients across Australian, European and United States hospitals. BMC Musculoskelet Disord 18:207. doi: 10.1186/s12891-017-1569-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Gundtoft PH, Pedersen AB, Varnum C, Overgaard S. 2017. Increased mortality after prosthetic joint infection in primary THA. Clin Orthop Relat Res 475:2623–2631. doi: 10.1007/s11999-017-5289-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Hozack WJ, Parvizi J. 2011. New definition for periprosthetic joint infection. J Arthroplasty 26:1135. doi: 10.1016/j.arth.2011.09.025. [DOI] [PubMed] [Google Scholar]
- 6.Parvizi J, Zmistowski B, Berbari EF, Bauer TW, Springer BD, Della Valle CJ, Garvin KL, Mont MA, Wongworawat MD, Zalavras CG. 2011. New definition for periprosthetic joint infection: from the Workgroup of the Musculoskeletal Infection Society. Clin Orthop Relat Res 469:2992–2994. doi: 10.1007/s11999-011-2102-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Osmon DR, Berbari EF, Berendt AR, Lew D, Zimmerli W, Steckelberg JM, Rao N, Hanssen A, Wilson WR, Infectious Diseases Society of America. 2013. Diagnosis and management of prosthetic joint infection: clinical practice guidelines by the Infectious Diseases Society of America. Clin Infect Dis 56:e1–e25. doi: 10.1093/cid/cis803. [DOI] [PubMed] [Google Scholar]
- 8.Peel TN, Dylla BL, Hughes JG, Lynch DT, Greenwood-Quaintance KE, Cheng AC, Mandrekar JN, Patel R. 2016. Improved diagnosis of prosthetic joint infection by culturing periprosthetic tissue specimens in blood culture bottles. mBio 7:e01776-15. doi: 10.1128/mBio.01776-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Tunney MM, Patrick S, Gorman SP, Nixon JR, Anderson N, Davis RI, Hanna D, Ramage G. 1998. Improved detection of infection in hip replacements. A currently underestimated problem. J Bone Joint Surg Br 80:568–572. [DOI] [PubMed] [Google Scholar]
- 10.Trampuz A, Piper KE, Jacobson MJ, Hanssen AD, Unni KK, Osmon DR, Mandrekar JN, Cockerill FR, Steckelberg JM, Greenleaf JF, Patel R. 2007. Sonication of removed hip and knee prostheses for diagnosis of infection. N Engl J Med 357:654–663. doi: 10.1056/NEJMoa061588. [DOI] [PubMed] [Google Scholar]
- 11.Piper KE, Jacobson MJ, Cofield RH, Sperling JW, Sanchez-Sotelo J, Osmon DR, McDowell A, Patrick S, Steckelberg JM, Mandrekar JN, Fernandez Sampedro M, Patel R. 2009. Microbiologic diagnosis of prosthetic shoulder infection by use of implant sonication. J Clin Microbiol 47:1878–1884. doi: 10.1128/JCM.01686-08. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Sampedro MF, Huddleston PM, Piper KE, Karau MJ, Dekutoski MB, Yaszemski MJ, Currier BL, Mandrekar JN, Osmon DR, McDowell A, Patrick S, Steckelberg JM, Patel R. 2010. A biofilm approach to detect bacteria on removed spinal implants. Spine 35:1218–1224. doi: 10.1097/BRS.0b013e3181c3b2f3. [DOI] [PubMed] [Google Scholar]
- 13.Janz V, Wassilew GI, Hasart O, Matziolis G, Tohtz S, Perka C. 2013. Evaluation of sonicate fluid cultures in comparison to histological analysis of the periprosthetic membrane for the detection of periprosthetic joint infection. Int Orthop 37:931–936. doi: 10.1007/s00264-013-1853-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Rothenberg AC, Wilson AE, Hayes JP, O'Malley MJ, Klatt BA. 2017. Sonication of arthroplasty implants improves accuracy of periprosthetic joint infection cultures. Clin Orthop Relat Res 475:1827–1836. doi: 10.1007/s11999-017-5315-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Tani S, Lepetsos P, Stylianakis A, Vlamis J, Birbas K, Kaklamanos I. 2017. Superiority of the sonication method against conventional periprosthetic tissue cultures for diagnosis of prosthetic joint infections. Eur J Orthop Surg Traumatol doi: 10.1007/s00590-017-2012-y. [DOI] [PubMed] [Google Scholar]
- 16.Yano MH, Klautau GB, da Silva CB, Nigro S, Avanzi O, Mercadante MT, Salles MJ. 2014. Improved diagnosis of infection associated with osteosynthesis by use of sonication of fracture fixation implants. J Clin Microbiol 52:4176–4182. doi: 10.1128/JCM.02140-14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Joseph L, Gyorkos TW, Coupal L. 1995. Bayesian estimation of disease prevalence and the parameters of diagnostic tests in the absence of a gold standard. Am J Epidemiol 141:263–272. doi: 10.1093/oxfordjournals.aje.a117428. [DOI] [PubMed] [Google Scholar]
- 18.Parvizi J, Gehrke T, International Consensus Group on Periprosthetic Joint Infections. 2014. Definition of periprosthetic joint infection. J Arthroplasty 29:1331. doi: 10.1016/j.arth.2014.03.009. [DOI] [PubMed] [Google Scholar]
- 19.Peel TN, Spelman T, Dylla BL, Hughes JG, Greenwood-Quaintance KE, Cheng AC, Mandrekar JN, Patel R. 2017. Optimal periprosthetic tissue specimen number for diagnosis of prosthetic joint infection. J Clin Microbiol 55:234–243. doi: 10.1128/JCM.01914-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Dendukuri NBP, Joseph L. 2014. BayesLatentClassModels: a program for estimating diagnostic test properties and disease prevalence, version 1.9.2, June 2013 McGill University, Montreal, Quebec, Canada. [Google Scholar]
- 21.Vergidis P, Greenwood-Quaintance KE, Sanchez-Sotelo J, Morrey BF, Steinmann SP, Karau MJ, Osmon DR, Mandrekar JN, Steckelberg JM, Patel R. 2011. Implant sonication for the diagnosis of prosthetic elbow infection. J Shoulder Elbow Surg 20:1275–1281. doi: 10.1016/j.jse.2011.06.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Minassian AM, Newnham R, Kalimeris E, Bejon P, Atkins BL, Bowler IC. 2014. Use of an automated blood culture system (BD BACTEC) for diagnosis of prosthetic joint infections: easy and fast. BMC Infect Dis 14:233. doi: 10.1186/1471-2334-14-233. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Islam K, Sayeed MA, Hossen E, Khanam F, Charles RC, Andrews J, Ryan ET, Qadri F. 2016. Comparison of the performance of the TPTest, Tubex, Typhidot and Widal immunodiagnostic assays and blood cultures in detecting patients with typhoid fever in Bangladesh, including using a Bayesian latent class modeling approach. PLoS Negl Trop Dis 10:e0004558. doi: 10.1371/journal.pntd.0004558. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Manning L, Laman M, Rosanas-Urgell A, Turlach B, Aipit S, Bona C, Warrell J, Siba P, Mueller I, Davis TM. 2012. Rapid antigen detection tests for malaria diagnosis in severely ill Papua New Guinean children: a comparative study using Bayesian latent class models. PLoS One 7:e48701. doi: 10.1371/journal.pone.0048701. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Tarr GA, Eickhoff JC, Koepke R, Hopfensperger DJ, Davis JP, Conway JH. 2013. Using a bayesian latent class model to evaluate the utility of investigating persons with negative polymerase chain reaction results for pertussis. Am J Epidemiol 178:309–318. doi: 10.1093/aje/kws465. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Peel TN, Sedarski JA, Dylla BL, Shannon SK, Amirahmadi F, Hughes JG, Cheng AC, Patel R. 2017. Laboratory workflow analysis of culture of periprosthetic tissues in blood culture bottles. J Clin Microbiol 55:2817–2826. doi: 10.1128/JCM.00652-17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Flayhart D, Borek AP, Wakefield T, Dick J, Carroll KC. 2007. Comparison of BACTEC PLUS blood culture media to BacT/Alert FA blood culture media for detection of bacterial pathogens in samples containing therapeutic levels of antibiotics. J Clin Microbiol 45:816–821. doi: 10.1128/JCM.02064-06. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Hughes HC, Newnham R, Athanasou N, Atkins BL, Bejon P, Bowler IC. 2011. Microbiological diagnosis of prosthetic joint infections: a prospective evaluation of four bacterial culture media in the routine laboratory. Clin Microbiol Infect 17:1528–1530. doi: 10.1111/j.1469-0691.2011.03597.x. [DOI] [PubMed] [Google Scholar]
- 29.Velay A, Schramm F, Gaudias J, Jaulhac B, Riegel P. 2010. Culture with BACTEC Peds Plus bottle compared with conventional media for the detection of bacteria in tissue samples from orthopedic surgery. Diagn Microbiol Infect Dis 68:83–85. doi: 10.1016/j.diagmicrobio.2010.04.010. [DOI] [PubMed] [Google Scholar]
- 30.Hughes JG, Vetter EA, Patel R, Schleck CD, Harmsen S, Turgeant LT, Cockerill FR III. 2001. Culture with BACTEC Peds Plus/F bottle compared with conventional methods for detection of bacteria in synovial fluid. J Clin Microbiol 39:4468–4471. doi: 10.1128/JCM.39.12.4468-4471.2001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Portillo ME, Salvado M, Trampuz A, Plasencia V, Rodriguez-Villasante M, Sorli L, Puig L, Horcajada JP. 2013. Sonication versus vortexing of implants for diagnosis of prosthetic joint infection. J Clin Microbiol 51:591–594. doi: 10.1128/JCM.02482-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Holinka J, Bauer L, Hirschl AM, Graninger W, Windhager R, Presterl E. 2011. Sonication cultures of explanted components as an add-on test to routinely conducted microbiological diagnostics improve pathogen detection. J Orthop Res 29:617–622. doi: 10.1002/jor.21286. [DOI] [PubMed] [Google Scholar]
- 33.Puig-Verdie L, Alentorn-Geli E, Gonzalez-Cuevas A, Sorli L, Salvado M, Alier A, Pelfort X, Portillo ME, Horcajada JP. 2013. Implant sonication increases the diagnostic accuracy of infection in patients with delayed, but not early, orthopaedic implant failure. Bone Joint J 95-B(2):244–249. doi: 10.1302/0301-620X.95B2.30486. [DOI] [PubMed] [Google Scholar]
- 34.Zitron R, Wajsfeld T, Klautau GB, da Silva CB, Nigro S, Mercadante MT, Polesello GC, Cury Rde P, Salles MJ. 2016. Concentration of sonication fluid through centrifugation is superior to membrane filtration for microbial diagnosis of orthopedic implant-associated infection. J Clin Microbiol 54:788–790. doi: 10.1128/JCM.02427-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Portillo ME, Salvado M, Alier A, Martinez S, Sorli L, Horcajada JP, Puig L. 2014. Advantages of sonication fluid culture for the diagnosis of prosthetic joint infection. J Infect 69:35–41. doi: 10.1016/j.jinf.2014.03.002. [DOI] [PubMed] [Google Scholar]
- 36.Marculescu CE, Berbari EF, Cockerill FR III, Osmon DR. 2006. Fungi, mycobacteria, zoonotic and other organisms in prosthetic joint infection. Clin Orthop Relat Res 451:64–72. doi: 10.1097/01.blo.0000229337.21653.f2. [DOI] [PubMed] [Google Scholar]
- 37.Berbari EF, Marculescu C, Sia I, Lahr BD, Hanssen AD, Steckelberg JM, Gullerud R, Osmon DR. 2007. Culture-negative prosthetic joint infection. Clin Infect Dis 45:1113–1119. doi: 10.1086/522184. [DOI] [PubMed] [Google Scholar]
- 38.Farrell JJ, Larson JA, Akeson JW, Lowery KS, Rounds MA, Sampath R, Bonomo RA, Patel R. 2014. Ureaplasma parvum prosthetic joint infection detected by PCR. J Clin Microbiol 52:2248–2250. doi: 10.1128/JCM.00432-14. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Thoendel M, Jeraldo P, Greenwood-Quaintance KE, Chia N, Abdel MP, Steckelberg JM, Osmon DR, Patel R. 2017. A novel prosthetic joint infection pathogen, Mycoplasma salivarium, identified by metagenomic shotgun sequencing. Clin Infect Dis 65:332–335. doi: 10.1093/cid/cix296. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.