Skip to main content
. 2018 Apr 21;81(6):911–921. doi: 10.1007/s10337-018-3520-6

Table 1.

Method validation results

Compound Sensitivity Linearity Expected concentration (µM) Precision and accuracy
Within-run (n = 5) Between-run (n = 10)
LOD (µM) LOQ (µM) Regression coefficient (R2) Slope CV (%)a Measured concentration (µM) Precision (CV %) Accuracy (%) Measured concentration (µM) Precision (CV %) Accuracy (%)
l-Arginine 1.7 3.2 0.998 3.20 5
25
100
200
4.867 ± 0.198
25.49 ± 0.97
99.17 ± 2.68
203.08 ± 4.70
9.14
3.86
1.64
2.32
102.74
98.09
100.83
98.49
5.142 ± 0.035
24.8 ± 01.01
101.85 ± 3.39
199.52 ± 5.44
3.63
4.09
3.33
2.73
97.23
100.87
98.19
100.24
ADMA 0.03 0.08 0.997 3.66 0.05
0.25
1.0
2.5
0.058 ± 0.0001
0.233 ± 0.0003
1.012 ± 0.0004
2.493 ± 0.008
14.88
7.53
2.09
3.52
86.10
107.14
98.80
100.30
0.056 ± 0.0003
0.235 ± 0.022
1.022 ± 0.049
2.477 ± 0.076
10.21
9.42
4.84
3.07
89.29
106.47
97.89
100.93
SDMA 0.02 0.05 0.997 3.33 0.05
0.25
1.0
2.5
0.058 ± 0.00003
0.239 ± 0,0002
0.990 ± 0,002
2.505 ± 0.006
8.99
6.39
3.98
3.20
85.62
104.53
101.03
99.81
0.057 ± 0.00003
0.230 ± 0.016
1.013 ± 0.038
2.506 ± 0.077
9.92
7.11
3.77
3.09
87.72
108.79
98.74
99.75
l-Citrulline 0.36 1.08 0.998 6.12 1
10
20
50
1.036 ± 0.003
10.00 ± 0.07
20.304 ± 0.428
50.019 ± 0.829
5.35
2.58
3.22
1.82
96.53
101.14
98.50
99.96
1.034 ± 0.093
9.88 ± 0.044
20.083 ± 0.617
50.163 ± 1.47
9.01
2.13
3.07
2.93
96.72
101.25
99.58
99.67
DMA 0.06 0.19 0.992 3.89 0.014
0.7
2.8
7.0
0.164 ± 0.00026
0.698 ± 0.001
2.71 ± 0.034
7.19 ± 0.016
9.84
5.49
6.80
1.75
85.32
100.26
103.18
97.30
0.152 ± 0.00004
0.64 ± 0.06
2.61 ± 0.15
7.46 ± 0.19
4.19
9.74
5.56
5.87
92.41
109.69
107.24
93.81

LOD limit of detection, LOQ limit of quantification, CV coefficient of variation