Table 2a.

Inflow orifice areas of the CoreValve defined in Fig. 3a at experimental room temperature and at 37°C.

Axial Positions (mm) IOA (cm2) STD
Calcified SAV ViV 37°C +6 3.3 0.15
0 2.2 0.12
−6.2 1.6 0.02
−9.8 1.4 0.06
Room Temperature +6 3.3 0.11
0 2.2 0.07
−6.2 1.6 0.09
−9.8 1.5 0.08
Non-Calcified SAV ViV 37°C +6 3.2 0.09
0 2.0 0.07
−6.2 1.7 0.08
−9.8 1.6 0.04
Room Temperature +6 3.1 0.22
0 2.0 0.09
−6.2 1.7 0.06
−9.8 1.5 0.08