Skip to main content
. 2018 Apr 17;19(3):360–366. doi: 10.1002/acm2.12328

Table 3.

Absorbed doses and CBCTDIw values obtained using 16‐cm‐diameter cylindrical phantom

SCBCT (tube 1) SCBCT (tube 2) DCBCT
Center 47.1 ± 0.2 50.9 ± 0.1 53.3 ± 0.1
Top 50.1 ± 0.1 96.5 ± 0.1 89.3 ± 0.1
Left 25.8 ± 0.1 51.4 ± 0.1 34.2 ± 0.1
Bottom 66.4 ± 0.1 29.6 ± 0.1 42.0 ± 0.1
Right 89.3 ± 0.1 77.0 ± 0.1 98.0 ± 0.1
CBCTDIw 54.3 ± 0.1 59.4 ± 0.1 61.7 ± 0.1

All values are expressed in mGy.

All values are average values ± standard deviation (SD).