Table 1.

Comparison between doses (cGy) calculated with Oncentra (air in balloon treated as water) and Monte Carlo (air in balloon modeled as air) for two patient cases of 35 mm Capri applicator

35 mm diameter case 1 35 mm diameter case 2
Point # TPS dose MC dose % difference Point # TPS dose MC dose % difference
1 441.6 462.7 4.8 1 395.3 411.2 4.7
2 483.6 500.9 3.6 2 476.9 496.0 4.0
3 334.7 354.9 6.1 3 291.3 311.7 7.0
4 479.8 498.2 3.8 4 479.1 496.6 3.7
5 423.5 444.6 5.0 5 403.0 420.6 4.4
6 499.0 516.1 3.4 6 498.4 517.2 3.8
7 358.7 380.8 6.2 7 302.8 323.2 6.7
8 512.7 533.3 4.0 8 536.0 553.2 3.2
9 424.6 446.2 5.1 9 390.4 407.2 4.3
10 521.5 538.0 3.2 10 510.2 529.1 3.7