. 2018 Mar 3;19(3):52–57. doi: 10.1002/acm2.12276

Table 2.

Measured imaging doses per fraction to the breast with daily MV‐MV and MV‐CBCT imaging using the maximum field size of 28 × 28 cm2

Breast of interest Relative position to treatment site MV‐MV (cGy) MV‐CBCT (cGy)
Low‐dose High‐quality Low‐dose High‐quality
Left Ipsilateral 2.15 4.32 4.82 9.71
Right 1.71 3.40 4.76 9.64
Left Contralateral 1.35 2.76 2.85 5.77
Right 0.41 0.83 2.83 5.72