Skip to main content
. 2018 Feb 28;19(3):94–100. doi: 10.1002/acm2.12292

Table 2.

Incidence of baseline shifts exceeding 2 mm, 3 mm, and 5 mm for the SI, LR, AP, and 3D directions

Time block 1 Time block 2 Time block 3
>2 mm
SI 21.2% 34.5% 47.6%
LR 0.0% 3.4% 4.8%
AP 4.5% 17.2% 26.2%
3D 27.3% 50.0% 66.7%
>3 mm
SI 6.1% 20.7% 31.0%
LR 0.0% 0.0% 2.4%
AP 0.0% 5.2% 7.1%
3D 7.6% 25.9% 38.1%
>5 mm
SI 1.5% 13.8% 16.7%
LR 0.0% 0.0% 0.0%
AP 0.0% 1.7% 2.4%
3D 3.0% 15.5% 19.0%