Table 3.

Image quality consistency thresholds for the planar radiographic modalities

kV MV
Warning (%) Action (%) Warning (%) Action (%)
f30(lp/mm) 1 3 f30(lp/mm) 2 4
f40(lp/mm) 1 3 f40(lp/mm) 1 3
f50(lp/mm) 1 3 f50(lp/mm) 2 4
Noise 5 10 Noise 1 2
CNR 3 6 CNR 2 4
Dose 3 6
kVp 1 2

Sample size of 30 measurements.