Skip to main content
. 2018 Mar 9;19(3):64–70. doi: 10.1002/acm2.12289

Table 4.

Image quality consistency thresholds for the small(S20) CBCT

Warning (%) Action (%)
High contrast spatial resolution(lp/mm)
f30 8 16
f40 5 11
f50 5 11
HU constancy
Lung(PMP) 5 9
Water(Poly) 3 6
Air 2 5
Uniformity
Mean 3 6
Sigma 5 10
Noise
Mean 3 6
Sigma 6 11

Sample size of 30 measurements.