Skip to main content
. 2018 Mar 9;19(3):64–70. doi: 10.1002/acm2.12289

Table 5.

Image quality consistency thresholds for the Medium (M20) CBCT

Warning (%) Action (%)
High contrast spatial resolution(lp/mm)
f30 9 18
f40 4 9
f50 4 8
HU constancy
Lung(PMP) 2 4
Water(Poly) 2 4
Air 1 3
Uniformity
Mean 1 3
Sigma 3 6
Noise
Mean 1 3
Sigma 4 8

Sample size of 30 measurements.