. 2018 Mar 9;19(3):64–70. doi: 10.1002/acm2.12289

Table 6.

Comparison of Image quality action thresholds for the planar radiographic modalities between the XVI and OBI

XVI (%) OBI* (%)
kV
High contrast spatial resolution(lp/mm)
f30 3 4
f40 3 3
f50 3 3
Dose 6 2
kVp 2 2
MV
High contrast spatial resolution(lp/mm)
f30 4 4
f40 3 3
f50 4 3

*Data from Stanley et al.7