Skip to main content
. 2018 Mar 9;19(3):64–70. doi: 10.1002/acm2.12289

Table 7.

Comparison of image quality action thresholds for the Small (S20) and Full‐Fan CBCT

XVI (%) OBI* (%)
High contrast spatial resolution(lp/mm)
f30 16 18
f40 11 18
f50 11 16
HU constancy
Lung(PMP) 9 6
Water(Poly) 6 12
Uniformity 10 12
Noise 11 12

*Data from Stanley et al.7