Skip to main content
. 2018 Mar 9;19(3):64–70. doi: 10.1002/acm2.12289

Table 8.

Comparison of image quality action thresholds for the Medium (M20) and Half‐Fan CBCT

XVI (%) OBI* (%)
High contrast spatial resolution(lp/mm)
f30 18 20
f40 9 22
f50 8 34
HU constancy
Lung(PMP) 4 8
Water(Poly) 4 12
Uniformity 6 17
Noise 8 16

*Data from Stanley et al.7