Published in final edited form as: NMR Biomed. 2018 Apr 19;31(6):e3921. doi: 10.1002/nbm.3921

Table 1.

The resonant frequency and quality factor of the reference coil are altered the same way by loading a mouse brain or the homogeneous phantom into the coil, indicating that SNR measurements on the homogenous phantom are representative of mouse brain imaging.

Resonant Frequency (MHz)

Unloaded Loaded with Brain Loaded with Phantom

Reference Coil 300.36 299.41 299.55
Quality Factor

Unloaded Loaded with Brain Loaded with Phantom

Reference Coil (Unshielded) 720 380 395
Coil 1 (Shielded) 705 370 385
Coil 2 (Shielded) 710 380 380

Also, the quality factor of each shielded coil in the two-coil system remains similar to the unshielded reference coil, demonstrating that the shield causes negligible RF losses, and that all three coils are almost identical.