Published in final edited form as: Magn Reson Med. 2016 Oct 7;78(1):383–386. doi: 10.1002/mrm.26344

Table 1.

Results of head coil former on SAR1g around the implanted lead tips and B1 field homogeneity on the mid-transverse slice in brain as well as in the entire cerebrum with the body coil driven to produce a 1μT B1+ field at the center of brain.

Relative permittivity of the head coil former Peak SAR/|B1+|2 W/kg/(μT)2 Coefficient of Variation of |B1+| on transverse plane Coefficient of Variation of |B1+| in cerebrum
1 0.1153 0.0763 0.1408
300 0.0717 0.0738 0.1092
400 0.0530 0.0697 0.1076
500 0.0392 0.0688 0.1053
600 0.0306 0.0665 0.0981
700 0.0229 0.0683 0.1041
900 0.0152 0.0692 0.1208