Intra-day and inter day accuracies and precisions of CUR (n=6)

Concentration (ng/ml) Intra-day mean (ng/ml) RSD (%) Accuracy (%) Inter-day Mean (ng/ml) RSD (%) Accuracy (%)
1 1.03 4.42 3.17 1.01 5.05 1.12
5 4.99 3.60 −0.12 5.00 3.60 0.09
50 50.42 3.83 0.85 49.53 3.32 −0.94
400 412.58 1.87 3.14 406.21 2.00 1.55
1200 1209.69 1.87 0.81 1210.87 2.39 0.91
3000 2846.70 1.87 −5.11 2945.93 2.26 −1.80