Table 5. Limit of detection for integrated microfluidics circuit (IFC) library preparation method.
IFC | BRAF_V600E | KRAS_G12V | EGFR E746_A750del | FLT3 D600_L601ins10 | JAK2_V617F | TP53_R273H | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Predicted | Observed | Predicted | Observed | Predicted | Observed | Predicted | Observed | Predicted | Observed | Predicted | Observed | ||||||
Straight | – | 0.150 | – | 0.571 | – | 0.757 | – | 0.115 | – | 0.533 | – | 0.372 | |||||
1:4 | 0.037 | 0.026 | 0.143 | 0.269 | 0.189 | 0.549 | 0.029 | 0.041 | 0.133 | 0.217 | 0.093 | 0.140 | |||||
1:16 | 0.009 | 0.009 | 0.036 | 0.085 | 0.047 | 0.327 | 0.007 | 0.004 | 0.033 | 0.095 | 0.023 | 0.048 | |||||
1:64 | 0.002 | 0.010 | 0.009 | 0.020 | 0.012 | 0.093 | 0.002 | 0.004 | 0.008 | 0.022 | 0.006 | 0.021 |