Table S2. Accuracy of VAF results with downsampling by amplicon to maximum 7,500× coverage.
Sample type | Variant (VAF) without downsampling | VAF downsampling cutoff =7,500 |
---|---|---|
Fixed | BRAF V600R (0.86; 2 bp indel) | 0.87 |
Unfixed | CALR L367Tfs*46 (0.44) | 0.47 |
Fixed | KRAS G12G (0.22) | 0.24 |
Unfixed | FLT3 ITD (0.05; 90 bp ins); NPM1 Y288Cfs (0.39) | 0.06; 0.40 |
Fixed | EGFR p.L747_T751del (0.39) | 0.37 |
Fixed | KIT Q556_K558delinsH (0.29) | 0.30 |
Fixed | NRAS Q61L (0.41) | 0.37 |