Skip to main content
. 2018 May;6(9):162. doi: 10.21037/atm.2018.05.07

Table S3. Intra-run replicate precision.

Sample_1 Sample_2 Concordant variants Discrepant variants Cause(s) of discrepancy
Integrated microfluidics circuit (IFC)
   BM1_DUP BM1 8 0
   PB1_DUP PB1 8 0
   CELL_DUP CELL 16 3 Intronic homopolymer [1];
single nucleotide variants (SNV) threshold [1];
indel mapping [1]
   FFPE1_DUP FFPE1 6 5 Intronic homopolymer [5]
   FFPE2_DUP FFPE2 6 1 Intronic homopolymer [1]
   FFPE3_DUP FFPE3 7 6 Intronic homopolymer [6]
   FFPE4_DUP FFPE4 6 5 Intronic homopolymer [5]
Low input
   FFPE9_DUP FFPE9 1 0
   FFPE10_DUP FFPE10 1 0
   FFPE11_DUP FFPE11 0 0
   FFPE12_DUP FFPE12 0 0
   FFPE13_DUP FFPE13 0 0
   FFPE14_DUP FFPE14 1 0
   FNA1_DUP FNA1 1 0
   FNA2_DUP FNA2 1 0
   FNA3_DUP FNA3 1 0
   FNA4_DUP FNA4 0 0
   FNA5_DUP FNA5 0 0
   FNA6_DUP FNA6 0 0