. 2018 May 5;7(3):226–231. doi: 10.1302/2046-3758.73.BJR-2017-0193.R1

Table II.

Results of the Bonferroni post hoc analysis. Significance was set as < 0.01. There was a significant difference in push-in force between the current interference and what we are investigating to be the optimal range of interference. There was no difference in the pull-out between the 0.5, 0.6 and 0.7 slot. In contrast, the interference values in the ‘optimal range’ showed a superior pull-out force compared with an interference of 1.1 mm and 1.3 mm

Push-in force
Current range
Interference 0.9 mm 1.1 mm 1.3 mm
0.5 mm < 0.01 < 0.01 < 0.01
Optimal range 0.6 mm < 0.01 < 0.01 < 0.01
0.7 mm < 0.01 < 0.01 < 0.01
Pull-out force
Current range
Interference 0.9 mm 1.1 mm 1.3 mm
0.5 mm 1 < 0.01 < 0.01
Optimal range 0.6 mm 1 < 0.01 < 0.01
0.7 mm 0.1 < 0.01 < 0.01