Table 1.

Parameters of the analytical procedure: recoveries for certified reference materials (CRMs, n = 10) with relative standard deviation (RSD) of 10 replicates

Metals Wavelength λ (nm) LoD (mg/L) LoQ (mg/L) Recovery (%) RSD (%)
Ca 422.7 0.0642 0.514 108.2 6.6
Cd 228.8 0.0621 0.010 91.6 4.3
Cu 324.8 0.0876 0.035 99.6 2.7
Fe 248.3 0.005 0.415 101.6 6.2
Hga 253.7 0.27 97.1 2.6
Mg 285.2 0.00015 0.017
Ni 232.00 0.006 0.013
Pb 217.0 0.1039 0.107 93.7 6.2
Zn 213.9 0.0170 0.024 106.2 5.3

For Cd and Pb, ERMCE195 CRM were used; for Hg, BCR-463 CRM was used and for other elements, SRM1577b; no appropriate CRM for Mg and Ni were found

LoD limits of detection (in solution), LoQ limits of quantification established for wine samples

aLoQ value for Hg expressed as a nanograms per liter