Table 1.

Aerosol Performance Characteristics of eFlow Closed System Nebulizers with Glycopyrrolate (N = 15, Mean ± Standard Deviation)

Glycopyrrolate strength Nebulization time (minutes) MMAD (μm) GSD FPD (μg) FPF (%) NGI recovery (% nominal dose)
25 μg/mL 1.88 ± 0.23 3.71 ± 0.26 1.66 ± 0.06 15.20 ± 1.26 72.28 ± 5.57 95.61 ± 1.90
50 μg/mL 1.92 ± 0.14 3.74 ± 0.20 1.65 ± 0.04 30.44 ± 1.51 71.65 ± 4.34 94.94 ± 1.75

CS, Closed System; FPD, fine particle dose; FPF, fine particle fraction; GSD, geometric standard deviation; MMAD, mass median aerodynamic diameter; NGI, Next Generation Impactor.