Skip to main content
. 2017 Aug 17;22(12):121604. doi: 10.1117/1.JBO.22.12.121604

Table 3.

Deviations over 2 years in broadband absorption and reduced scattering of a Ph-A phantom measured at UCI by four different operators, test B1.

Test B1 μa (mm1) μs (mm1)
Stdev NSD (%) Stdev NSD (%)
Spectral average 0.0004 5.3 0.02 1.8
Spectral min 0.0002 2.2 0.01 1.3
Spectral max 0.0007 8.9 0.03 2.5
Spectral median 0.0003 5.7 0.02 1.7