Skip to main content
. 2018 Mar 25;56(6):546–558. doi: 10.1002/mrc.4717

Figure 8.

Figure 8

Testing the stability of a newly installed Bruker Neo 500 MHz NMR spectrometer. 480 1H NMR experiments were acquired over 8 hr using a sample of acetone in D2O. The GNAT analysis tool is used to monitor changes in the linewidth (top) and phase (bottom) of the water signal as a function of time. The effects of room temperature variation are apparent