Table 5.
Comparison of Y-90 PET/CT vs. 99mTc-MAA imaging-based LSFs for the 20 patients that had post-therapy imaging
LSF, mean (range) | P * | Maximum absolute difference in LSF (%) | |
---|---|---|---|
99mTc-MAA planar imaging | 4.1% (1.2–15.0) | 0.0002 | 12.3 |
99mTc-MAA SPECT/CT (AC and SC) | 1.0% (0.4–1.6) | 0.968 | 1.6 |
90Y PET/CT | 1.0% (0.3–2.8) |
*P value for significance of difference between mean LSF from indicated method and Y-90 PET/CT