Table 2. LDCT Scanners and Scanning Parameters Used in Our Pilot Study for K-LUCAS Project.

Total (n = 256) GE Discovery 750 (n = 32) GE Revolution (n = 9) Somatom Definition (n = 101) Somatom Definition Flash (n = 22) Brilliance 64 (n = 92)
Scanning parameters
 Scan mode Spiral Spiral Spiral Spiral Spiral
 Rotation time (s) 0.5 0.5 0.33 0.28 0.5
 Collimation (mm) 0.625 × 64 0.625 × 64 0.6 × 32 × 2 0.6 × 64 × 2 0.625 × 64
 Pitch 0.984 0.984 1 1 1
 kVp 120 (n = 3) 100 120 120 120
100 (n = 29)
 mAs 50 30 30 (n = 1) 25 (n = 18)
50 (n = 4)
 Reference mAs 25 (n = 100) 26
Reconstruction
 Slice thickness/interval (mm) 1.25/1.25 for axial images 1.25/1.25 for axial images 1/1 for axial images 1/1 for axial images 1/1 for axial images
2.5/2.5 for axial and coronal images 2.5/2.5 for axial and coronal images 3/3 for axial and coronal images 2.5/2.5 for axial images
 Reconstruction algorithm Lung for 1.25-mm images Lung for 1.25-mm images B50 for 1-mm images I70f for 1-mm images YA for 1-mm images
Standard for 2.5-mm images Standard for 2.5-mm images B31 for 3-mm images I50f for 3-mm images A for 2.5-mm images
 Iterative reconstruction ASIR 50 ASIR-V 50 SAFIRE level 2 SAFIRE level 1

ASIR = adaptive statistical iterative reconstruction, LDCT = low-dose chest computed tomography, SAFIRE = Sinogram Affirmed Iterative Reconstruction