Table 2.

Diagnostic Accuracy of Cardiio Rhythm Facial and Fingertip Photoplethysmography in Detecting AF

Variable AF Present on 12‐Lead ECG AF Absent on 12‐Lead ECG Total
Cardiio Rhythm facial photoplethysmography positive 71 (32.7) 6 (2.8) 77
Cardiio Rhythm facial photoplethysmography negative 4 (1.8) 136 (62.7) 140
Total 75 142 217
Cardiio Rhythm fingertip photoplethysmography positive 71 (32.7) 10 (4.6) 81
Cardiio Rhythm fingertip photoplethysmography negative 4 (1.8) 132 (60.8) 136
Total 75 142 217

Data are given as number (percentage). The percentages were the number in each cell divided by the total number of tests.

AF indicates atrial fibrillation.