Table 4.
Radiopharmaceuticals | RCP-1 (%) 1 | RCP-2 (%) 2 | RCP-3 (%) 3 | RCP (%) Expected Value |
---|---|---|---|---|
Ceretec (GE Healthcare) [99mTc]Tc-HMPAO N = 24 |
98.1 (±1.0) 31.0 °C (±0.5) |
94.8 (±1.1) 31.0 °C (±0.5) |
94.5 (±2.1) 31.0 °C (±0.5) |
≥80% |
MAASOL (GE Healthcare) [99mTc]Tc-albumin colloid N = 24 |
100.0 (±0.0) 32.0 °C (±0.5) |
100.0 (±0.0) 32.0 °C (±0.5) |
100.0 (±0.0) 30.0 °C (±0.5) |
≥95% |
Nanocoll (GE Healthcare) [99mTc]Tc-nanocolloid N = 80 |
99.10 (±0.09) 29.0 °C (±0.3) |
100.0 (±0.0) 29.0 °C (±0.3) |
99.3 (±1.2) 29.0 °C (±0.3) |
≥95% |
Renocis (IBA molecular) [99mTc]Tc-DMSA N = 8 |
100.0 (±0.9) 33.0 °C (±0.9) |
100.0 (±0.9) 29.0 °C (±0.9) |
100.0 (±0.9) 29.0 °C (±0.9) |
≥95% |
Technemibi (Mallinckrodt) [99mTc]Tc-MIBI N = 24 |
98.4 (±0.2) 29.0 °C (±0.5) |
99.1 (±0.7) 29.0 °C (±0.5) |
98.9 (±0.8) 29.0 °C (±0.5) |
≥94% |
Technescan HDP (Mallinckrodt) [99mTc]Tc-HDP N = 60 |
100.0 (±1.2) 33.0 °C (±0.3) |
98.8 (±1.6) 33.0 °C (±0.3) |
98.9 (±1.8) 31.0 °C (±0.3) |
≥95% |
Technescan MAG3 (Mallinckrodt) [99mTc]Tc-MAG3 N = 24 |
100.0 (±0.0) 30.0 °C (±0.5) |
99.5 (±0.5) 30.0 °C (±0.5) |
99.4 (±0.5) 29.0 °C (±0.5) |
≥94% |
1 RCP-1 = value at the end of the labelling and temperature at the time of test; 2 RCP-2 = value determined in the middle between the labelling and the expiry time and temperature at the time of test; 3 RCP-3 = value determined at the expiry time and temperature at the time of test; expected value is the standard value required by the manufacturer for administration to patients.