Skip to main content
. 2017 Jun 1;82(3):397–406. doi: 10.1093/neuros/nyx216

TABLE 4.

Subarachnoid Pressure Test

Baseline (mm Hg) Pmax (mm Hg)
Time point n Tachosil Hemopatch Tachosil Hemopatch P Value
Overall 18 3.23 (1.50) 4.21 (3.06) 13.9 (6.74) 12.3 (3.47) .46
Day 0 9 2.73 (1.23) 3.75 (2.19) 13.8 (8.52) 12.9 (4.41) .85
Day 14 9 3.74 (1.64) 4.67 (3.81) 14.0 (4.89) 11.7 (2.32) .39

TachoSil, a fibrinogen and thrombin-coated collagen pad, and Hemopatch, a polyethylene glycol-coated collagen dural sealant and substitute, withstood similar subarachnoid pressure prior to failure. Data presented as mean (standard deviation).