. 2018 May 18;8(2):50. doi: 10.3390/bios8020050

Figure 4.

Figure 4

Impact of the number of paired points on the uncertainty of MARD for an illustrative CGM system (CGM2, the GuardianTM REAL-Time unit): upper and lower bounds of the confidence interval (CI) with probability γ=0.95. The constant line represents the value to which it would converge for Nref towards infinity.