Skip to main content
. 2018 May 1;8(1):38–41. doi: 10.5005/jp-journals-10018-1255

Table 1: Mean of Ct values for each PCR instrument

     ABI 7500 FAST      LC 480-II      LC96      Mx3005p      PikoReal      Mean Ct      SD      CV (%)     
900,000 c/r      15.40      15.55      16.90      15.81      14.49      15.63      0.87      5.54     
90,000 c/r      18.94      18.28      20.51      18.59      17.34      18.73      1.16      6.18     
9,000 c/r      22.71      21.93      24.27      22.77      20.84      22.50      1.26      5.59     
900 c/r      26.51      25.2      27.73      26.45      24.18      26.01      1.36      5.23     
90 c/r      29.55      27.81      31.25      30.03      26.95      29.12      1.73      5.94     
9 c/r      33.00      29.62      35.37      32.72      28.40      31.82      2.80      8.80     
0.9 c/r      34.82      29.96      36.86      35.38      29.18      33.24      3.44      10.36     
0.09 c/r      39.10      30.96      38.73      42.33      29.70      36.16      5.52      15.28     

SD: Standard deviation