Table 4: Performance of the different primer suppliers (CIGB and BCD) on the Roche LightCycler® 480 II
| Mean Cp | SD | CV (%) | |||||||||||
| Concentration | CIGB | BCD | CIGB | BCD | CIGB | BCD | |||||||
| 900,000 c/r | 15.55 | 15.59 | 0.05 | 0.07 | 0.34 | 0.42 | |||||||
| 90,000 c/r | 18.28 | 18.43 | 0.05 | 0.14 | 0.28 | 0.75 | |||||||
| 9,000 c/r | 21.93 | 21.89 | 0.06 | 0.03 | 0.28 | 0.14 | |||||||
| 900 c/r | 25.2 | 25.14 | 0.12 | 0.14 | 0.46 | 0.55 | |||||||
| 90 c/r | 27.81 | 27.94 | 0.08 | 0.14 | 0.29 | 0.51 | |||||||
| 9 c/r | 29.62 | 29.51 | 0.12 | 0.27 | 0.39 | 0.92 | |||||||
| 0.9 c/r | 29.96 | 29.96 | 0.20 | 0.04 | 0.65 | 0.13 | |||||||
| 0.09 c/r | 30.96 | 30.55 | 0.49 | 0.04 | 1.58 | 0.14 | |||||||
Dilution series with the pST012012 plasmid at concentrations of 900,000 to 0.09 c/r were used, and three replicates per dilution step were measured in the qPCR. The means of Ct or Cp values were compared between the thermocyclers. SD: Standard deviation; CV (%): Coefficient of variation expressed in percent