Table 4: Performance of the different primer suppliers (CIGB and BCD) on the Roche LightCycler® 480 II

     Mean Cp      SD      CV (%)     
Concentration      CIGB      BCD      CIGB      BCD      CIGB      BCD     
900,000 c/r      15.55      15.59      0.05      0.07      0.34      0.42     
90,000 c/r      18.28      18.43      0.05      0.14      0.28      0.75     
9,000 c/r      21.93      21.89      0.06      0.03      0.28      0.14     
900 c/r      25.2      25.14      0.12      0.14      0.46      0.55     
90 c/r      27.81      27.94      0.08      0.14      0.29      0.51     
9 c/r      29.62      29.51      0.12      0.27      0.39      0.92     
0.9 c/r      29.96      29.96      0.20      0.04      0.65      0.13     
0.09 c/r      30.96      30.55      0.49      0.04      1.58      0.14     

Dilution series with the pST012012 plasmid at concentrations of 900,000 to 0.09 c/r were used, and three replicates per dilution step were measured in the qPCR. The means of Ct or Cp values were compared between the thermocyclers. SD: Standard deviation; CV (%): Coefficient of variation expressed in percent