Skip to main content
. 2018 Jun 17;4:2333721418781126. doi: 10.1177/2333721418781126

Table 4.

Percentage Accuracy of Activity Monitors Stratified by Gait Speed Category.

Activity monitor <0.6 m/s
M (SD)
n = 7
0.60-0.79 m/s
M (SD)
n = 12
0.8-1.0 m/s
M (SD)
n = 11
>1.0 m/s
M (SD)
n = 13
p value
Actigraph GT3X 14.1 (18.6) 35.6 (31.2) 52.7 (28.6) 85.1 (20.5) <.0001
ActivPAL 86.8 (14.0) 91.8 (15.1) 98.2 (0.8) 95.1 (9.2) .0174
StepWatch 99.4 (0.6) 98.2 (2.5) 99.7 (0.3) 99.0 (0.9) .0208

Note. Accuracy defined as monitor steps/observed steps × 100 with 100% indicating perfect accuracy.