. 2017 Dec 8;28(8):3141–3150. doi: 10.1007/s00330-017-5169-1

Table 1.

Sequences in MRI protocol

Parameter Axial T2 2D FSE DWI DKI MTI
TR (ms) 3000-5000 4000 6000 24
TE (ms) 99-106 70–75 94 2.4 / 4.8
Averages 3 8 6 1
Section thickness (mm) 3 4 3.6 4
Section gap (mm) 1 0 0.4 0
FOV (mm) 220 × 220 280 × 280 280 × 280 220 × 220
Matrix 384 × 288 128 × 128 128 × 96 192 × 160
Resolution (mm2) 0.6 × 0.8 2.2 × 2.2 2.2 × 2.9 1.15 × 1.38
Receiver bandwidth (± kHz) 50 111 111 31
ASSET factor No ASSET 2 2 No ASSET
Time 4:39 2:58 11:30 2:50
Other ETL 16
No phase wrap
b-values 0, 1400 s/mm2 b-values 150, 450, 800, 1150, 1500 s/mm2 Flip angle 5°

FSE, fast spin echo; EPI, echo planar imaging; DWI, diffusion-weighted imaging; FOV, field of view; ETL, echo train length; DKI, diffusion kurtosis imaging; MTI, magnetisation transfer imaging