Skip to main content
. 2018 Jul 4;6(7):E865–E871. doi: 10.1055/a-0599-5886

Table 1. Specific data on OPD and OFD.

Data Open-pore film drainage (OFD) Open-pore polyurethane foam drainage (OPD)
Material Very thin double layered film membrane Polyurethane foam
Diameter of drainage tube 4 – 6 mm 4 – 6 mm
Diameter of distal end of drainage tube with open-porosity element 4 – 6 mm 1.5 – 3.2 cm
Length of distal end of drainage tube with open porosity element 3 – 25 cm 3 – 12 cm
Visible effect of negative pressure Pimpled pattern Erosive pattern
Endoscopic techniques Push, pull, pull-through, guide-wire, Seldinger technique Push, pull, pull-through
Negative pressure therapy variants Intraluminal and intracavitary Intraluminal and intracavitary
Insertion Nasally and through openings smaller than 1.5 cm in diameter Orally and through openings more than 1.5 cm in diameter