Table 4.

Analytical characteristics of the detection of hydrogen peroxide in solutions with Au-, Pd-, and Pd-Au nanodendrite electrodes electrodeposited by DENA, 0.1 M PBS.

Metal/alloy Linear range, M LOD, M Sensitivity, A M−1
Au 2 × 10−4 − 1 × 10−2 5.9 × 10−5 5.7 × 10−7
Pd 1 × 10−5 − 7.9 × 10−3 6.0 × 10−6 5.4 × 10−7
Pd-Au 1 × 10−6 − 1 × 10−3 2.4 × 10−7 1.8 × 10−5