Skip to main content
. 2018 May 8;19(4):35–43. doi: 10.1002/acm2.12321

Table 3.

Target reproducibility results

Patient kV stability (mm) Clip migration (mm) TLE_BH (mm) TRE_BH (mm)
P1 0.8 ± 0.6 2.2 ± 1.4 4.7 ± 1.9 1.9 ± 1.3
P2 0.9 ± 0.9 1.9 ± 2.2 3.5 ± 3.1 2.6 ± 0.7
P3 1.2 ± 1.5 1.0 ± 1.1 8.4 ± 2.4 4.2 ± 1.1
P4 0.5 ± 0.4 2.0 ± 2.8 5.7 ± 4.3 3.8 ± 1.1
P5 0.7 ± 0.8 1.6 ± 2.5 8.7 ± 2.7 3.5 ± 1.0
P6 1.0 ± 1.0 1.0 ± 0.9 4.4 ± 1.2 2.9 ± 1.5
P7 0.8 ± 0.6 1.0 ± 1.4 7.1 ± 2.0 4.1 ± 1.4
P8 0.7 ± 0.8 1.3 ± 1.0 4.2 ± 1.8 3.0 ± 0.5
P9 0.7 ± 0.8 0.4 ± 0.5 5.8 ± 2.3 3.7 ± 3.2
P10 1.3 ± 0.7 1.8 ± 1.9 6.6 ± 2.1 3.8 ± 1.7
Median 0.9 ± 0.8 1.2 ± 1.5 5.8 ± 3.4 3.4 ± 1.7