Table 4.
Intra‐DIBH stability and intrafraction reproducibility from 31 left‐sided patients in this analysis compared to those of spirometry‐based results from seven patients as reported on in Fassi et al
| Intra‐DIBH variability (mm) | Intrafraction reproducibility (mm) | |||||
|---|---|---|---|---|---|---|
| A/P | S/I | R/L | A/P | S/I | R/L | |
| Fassi et al. (n = 7) Surface monitoring + spirometry | 1.37 | 1.78 | 0.74 | 2.16 | 2.30 | 1.88 |
| This analysis (n = 31) SI + voluntary DIBH | 0.66 | 0.58 | 0.49 | 2.17 | 1.98 | 1.52 |
| % within 2 mm | 72.1 | 76.3 | 84.6 | – | – | – |
| % within 3 mm | 98.2 | 98.8 | 99.6 | 74.6 | 76.0 | 88.4 |
| % within 4 mm | 99.4 | 99.6 | 100.0 | 88.1 | 89.5 | 95.7 |
| % within 5 mm | – | – | – | 94.9 | 96.5 | 98.2 |
| % within 7 mm | – | – | – | 99.2 | 99.5 | 99.6 |