Skip to main content
. 2018 Jun 1;19(4):163–172. doi: 10.1002/acm2.12360

Table 2.

Uncertainty analysis for MC simulations used in this study

Type A Type B
Statistical variation in absorbed dose determination from repeated MC simulation runs 1%
Variations of the source geometry from one source to another in manufacturing process24 0.5%
Uncertainty in cross‐section library data for Ir‐19244 0.5%
Effect of volume averaging on absorbed dose calculation in sensitive volumes45 0.1%
Uncertainty in composition of tissues used during source tracking simulations46, a 2%
Total TG‐43 simulation uncertainty 1.2%
Total source tracking simulation uncertainty 2.3%
a

Applies to source tracking simulations only.