Table 2.
Type A | Type B | |
---|---|---|
Statistical variation in absorbed dose determination from repeated MC simulation runs | 1% | |
Variations of the source geometry from one source to another in manufacturing process24 | 0.5% | |
Uncertainty in cross‐section library data for Ir‐19244 | 0.5% | |
Effect of volume averaging on absorbed dose calculation in sensitive volumes45 | 0.1% | |
Uncertainty in composition of tissues used during source tracking simulations46, a | 2% | |
Total TG‐43 simulation uncertainty | 1.2% | |
Total source tracking simulation uncertainty | 2.3% |
Applies to source tracking simulations only.