Skip to main content
. 2018 Jun 7;19(4):125–133. doi: 10.1002/acm2.12348

Table 1.

Rigid registration tests — translations only. The data are combined from Tables 5 and 6 in the Report

Case Stationary dataset Moving dataset Known shifts Known T (x,y,z) (mm)
1 Basic Phantom Dataset 2 (CT) Basic Phantom Dataset 1 (CT) Dataset 2 is shifted wrt Dataset 1 by 10 mm to patient Lt, 5 mm Ant, 15 mm Sup. (−10, 5, −15)
2 Basic Phantom Dataset 1 (PET)
3 Basic Phantom Dataset 1 (MR1)
4 Basic Phantom Dataset 1 (MR2)
5 Basic Phantom Dataset 1 (CBCT)
6 Basic Anatomical Dataset 1 (CT) Basic Anatomical Dataset 2 (CT) Datasets 2,3,4,5,6 shifted wrt Dataset 1 by 3 mm Lt, 5 mm Ant, 12 mm Sup. (3, −5, 12)
7 Basic Anatomical Dataset 3 (PET)
8 Basic Anatomical Dataset 4 (MRT1)
9 Basic Anatomical Dataset 5 (MRT2)