. 2017 Mar 31;3:5. doi: 10.1186/s41205-017-0013-4

Table 2.

Artifact Volume and segmentation time for 4 patients referred for head-and-neck acquisition and reconstructed with and without iterative metal artifact reconstruction. Only one patient was acquired with a jaw spacer, whereas none of the acquisitions were performed with a tilted gantry

Patient Iterative MAR Artifact Volume [ml] Segmentation time [s] Jaw spacer
1 No 36 340 No
Yes 5.6 200
2 No 4.5 150 No
Yes 4.5 135
3 No 36 300 No
Yes 3.3 180
4 No 6.7 310 Yes
Yes 3.4 101