Skip to main content
. 2018 Jun 26;25(Pt 4):1113–1122. doi: 10.1107/S1600577518007907

Table 2. Comparison of the experimental parameters for both capillaries.

Q: total volume flow rate; v: linear flow speed; t dwell: average dwell time in the X-ray beam; Re: Reynolds number (for water); N SAXS: number of SAXS curves taken; t tot: total exposure time of the sample. Due to the smaller diameter of the d i = 0.9 mm capillary, the flow speeds for equivalent flow rate are higher and the average dwell times shorter.

  v (mm s−1) t dwell (ms) Re    
Q (µl s−1) 1.7 mm 0.9 mm 1.7 mm 0.9 mm 1.7 mm 0.9 mm N SAXS t tot (ms)
0 0 0 1000 1000 0 0 20 1000
10 4.4 15.7 45 13 6 11 40 2000
20 8.8 31.4 23 6 11 22 20 1000
40 17.6 62.9 11 3 23 43 10 500