Skip to main content
. 2018 Jun 15;28(2):020709. doi: 10.11613/BM.2018.020709

Table 2. Internal quality control data for the analytes tested.

Analytes Reagent manufacturers Mean* SD CV
(%)
Mean
CV (%)
ALT
(U/L)
Abbott 28 0.93 3.37 2.70
97 1.96 2.03
Leadman 28 1.04 3.67 3.25
94 2.68 2.83
Mindray 29 0.86 3.02 2.63
97 2.18 2.24
Zybio 33 1.63 4.98 3.84
101 2.71 2.69
AST
(U/L)
Abbott 42 0.85 2.01 1.66
189 2.47 1.31
Leadman 41 1.13 2.74 2.24
190 3.28 1.73
Mindray 44 0.96 2.19 1.82
199 2.88 1.44
Zybio 44 1.61 3.61 2.40
195 2.33 1.19
urea
(mmol/L)
Abbott 5.5 0.12 2.21 2.41
17.1 0.44 2.6
Leadman 5.6 0.26 4.63 4.63
15.6 0.72 4.62
Mindray 6.0 0.14 2.37 3.02
16.5 0.06 3.67
Zybio 5.9 0.41 7.01 5.18
16.7 0.56 3.34
UA
(μmol/L)
Abbott 286 2.98 1.04 0.82
586 3.53 0.6
Leadman 289 2.73 0.95 0.87
572 4.45 0.78
Mindray 289 4.33 1.50 1.14
577 4.45 0.77
SD – standard deviation. CV – coefficient of variation. ALT – alanine aminotransferase. AST – aspartate aminotransferase. UA – uric acid. *The mean data of internal quality control with two levels were accumulated for 20 consecutive days. The mean CV were the average data of 2 QC levels of CV.