Skip to main content
. 2018 Jun 28;5:697–704. doi: 10.1016/j.mex.2018.06.017

Table 6.

MS/MS acquisition parameters.

MRM transitions
Analytes Q1 > Q3 RT (min) Dwell (s) Cone (V) CE (eV) Delay (s)
PFBAa 212.9 > 168.9 1.3 0.20 18.0 9.0 0.01
PFPeAa 262.8 > 219.1 1.6 0.20 16.0 9.0 0.01
PFHxA1a 312.7 > 269.0 2.5 0.10 15.0 9.0 0.01
PFHxA2 312.7 > 118.9 2.5 0.05 15.0 21.0 0.01
PFOA1a 412.6 > 369.0 3.7 0.10 18.0 10.0 0.01
PFOA2 412.6 > 169.2 3.7 0.05 18.0 18.0 0.01
PFBS1a 298.7 > 79.8 2.9 0.10 45.0 29.0 0.01
PFBS2 298.7 > 98.8 2.9 0.05 45.0 29.0 0.01
PFHxS1a 398.6 > 79.8 4.0 0.10 50.0 35.0 0.01
PFHxS2 398.6 > 98.8 4.0 0.05 50.0 30.0 0.01
PFOS1a 498.5 > 79.8 4.9 0.10 60.0 45.0 0.01
PFOS2 498.5 > 98.9 4.9 0.10 60.0 40.0 0.01
MPFBAb 216.9 > 172.1 1.3 0.10 15.0 10.0 0.01
MPFPeAb 267.8 > 223.1 1.6 0.20 15.0 9.0 0.01
MPFHxAb 314.6 > 270.0 2.5 0.10 15.0 10.0 0.01
MPFOAb 416.7 > 371.9 3.7 0.10 15.0 11.0 0.01
MPFHxSb 402.6 > 83.8 4.0 0.10 55.0 35.0 0.01
MPFOSb 502.5 > 79.9 4.9 0.10 60.0 40.0 0.01
a

Primary transition used for quantitation. Secondary transitions (when available) used for confirmation.

b

Stable isotope label internal standards.