Table 1. Protein, latex allergen, and single allergen content of gloves (G) after extraction according to the standard EN 455-3.

Product Name Protein content (µg/g) Latex allergen content (µg/g) Hev b 1 content (ng/g) Hev b 3 content (ng/g) Hev b 5 content (ng/g) Hev b 6.02 content (ng/g) Sum of single allergens (ng/g)
G01 Negative control 73.4 ND ND ND ND ND ND
G02 Positive control 617.4 361.7 7,554.4 ND 18,521.0 96,623.0 122,698.4
G03 Comfort 35.6 ND ND ND ND ND ND
G04 Contact 53.1 15.0 2,566.0 2,271.3 997.2 ND 5,834.5
G05 Derma Skin 22.0 1.4 96.4 ND 104.8 ND 201.2
G06 Micro-Thin Nutex 35.1 ND 471.3 488.6 ND ND 959.9
G07 Micro-Touch 28.6 1.8 72.4 ND ND 246.0 318.4
G08 Biogel Super Sensitive 11.9 1.2 43.3 ND ND 55.4 98.7
G09/1 Biogel Eclipse Indicator 22.1 1.0 65.2 ND ND 88.5 153.7
G09/2 Biogel Eclipse Indicator 19.9 1.6 238.2 ND ND 197.0 435.2
G10 Peha-soft unsterile ND ND 229.5 ND 34.2 ND 263.8
G11 Peha-soft steril ND 0.9 221.5 ND 50.2 ND 271.7
G12 Peha-micron plus 32.8 11.3 917.6 688.0 979.5 442.8 3,027.9
G13 Peha-taft plus 13.4 2.7 125.3 63.1 78.8 74.1 341.3
G14 Gentle Skin classic ND ND ND ND ND ND ND
G15 Gentle Skin Anatom 7.1 ND ND ND ND ND ND
G16 Gentle Skin grip 92.3 11.6 950.3 177.5 580.6 1,936.8 3,645.2
G17 Augustus polymer ND 1.3 54.0 ND 68.7 ND 122.8
G18 Augustus puderfrei 17.1 2.1 41.3 ND 346.9 ND 388.2
G19 Augustus-Gel ND ND ND ND 39.1 ND 39.1

Given are the mean values of the single measurements. The determination of the protein content is based on 4 single measurements, that of the latex allergen and the Hev b 1 contents as double measurements, and the determination of the single allergens Hev b 3, 5, and 6.02 as a single measurement. The negative control (G01) could only be extracted with the IPA method and not with the EN method. If all single measurements were below the detection limit, the content is indicated by ND (not detectable).